Strona 1 - OPERATION MA

OPERATION MANUAL DAKOTA ULTRASONICS MODEL PR-82 SONIC TESTER P/N P-102-0004 Rev 1.90, January 2008

Strona 2

Dakota Ultrasonics 6The Display The numeric portion of the display consists of 4 complete digits preceded by a leading "1", and is used to

Strona 3 - DISCLAIMER

PR-82 Sonic Tester 7 When the IN symbol is on, the PR-82 is displaying a thickness value in inches. The maximum thickness that can be displayed is 1

Strona 4

Dakota Ultrasonics 8The Transducer The transducer is the "business end" of the PR-82. It transmits and receives the ultrasonic sound waves

Strona 5 - INTRODUCTION

PR-82 Sonic Tester 9 This is a top view of a typical transducer. Press against the top with the thumb or index finger to hold the transducer in plac

Strona 6

Dakota Ultrasonics 10If the Stability Indicator has fewer than five bars darkened, or the numbers on the display seem erratic, first check to make su

Strona 7 - OPERATION

PR-82 Sonic Tester 11Condition and Preparation of Surfaces In any ultrasonic measurement scenario, the shape and roughness of the test surface are of

Strona 8

Dakota Ultrasonics 12Probe Zero Setting the Zero Point of the PR-82 is important for the same reason that setting the zero on a mechanical micrometer

Strona 9

PR-82 Sonic Tester 13At this point, the PR-82 has successfully calculated it's internal error factor, and will compensate for this value in any

Strona 10 - The Display

Dakota Ultrasonics 14Calibration to a known thickness NOTE: This procedure requires a sample piece of the specific material to be measured, the exac

Strona 11 - Sonic Tester

PR-82 Sonic Tester 15Calibration to a known velocity NOTE: This procedure requires that the operator know the sound-velocity of the material to be m

Strona 12 - The Transducer

Copyright  2008 Dakota Ultrasonics. All rights reserved. No part of this publication may be reproduced, translated into another language, stored in

Strona 13 - Making Measurements

Dakota Ultrasonics 16Two Point Calibration NOTE: This procedure requires that the operator has two known thickness points on the test piece that ar

Strona 14 - IMPORTANT

PR-82 Sonic Tester 17Scan Mode While the PR-82 excels at making single point measurements, it is sometimes desirable to examine a larger region, sear

Strona 15

Dakota Ultrasonics 18 TRANSDUCER SELECTION The PR-82 is inherently capable of performing measurements on a wide range of materials, from various meta

Strona 16 - Performing a Probe-Zero

PR-82 Sonic Tester 19strength of the waves, and thus, the PR-82's ability to detect the returning echo. Higher frequency ultrasound is absorbed

Strona 17 - Calibration

Dakota Ultrasonics 20Selection of the proper transducer is often a matter of tradeoffs between various characteristics. It may be necessary to exper

Strona 18 - Dakota Ultrasonics

PR-82 Sonic Tester 21 APPENDIX A Product Specifications Physical Weight: 10 ounces Size: 2.5W x 4.75H x 1.25D inches (63.5W x 120.7H x 31.8D

Strona 19

Dakota Ultrasonics 22 blank page

Strona 20 - Two Point Calibration

PR-82 Sonic Tester 23 APPENDIX B Application Notes • Measuring pipe and tubing When measuring a piece of pipe to determine the thickness of the pipe

Strona 21 - Scan Mode

Dakota Ultrasonics 24above this point, the change in sound velocity of the material being measured starts to have a noticeable effect upon ultrasonic

Strona 22 - TRANSDUCER SELECTION

PR-82 Sonic Tester 25An additional important consideration when measuring laminates, is that any included air gaps or pockets will cause an early ref

Strona 23

PR-8 2 Ultrasonic Thickness Gauge CONTENTS INTRODUCTION OPERATION THE KEYPAD THE DISPLAY THE TRANSDUCER MAKING MEASUREMENTS CONDITION AND PREPARATION

Strona 24

Dakota Ultrasonics 26 blank page

Strona 25 - APPENDIX A

PR-82 Sonic Tester 27 APPENDIX C Sound Velocities of some Common Materials Material sound velocityin/usm/sAluminum0.2506350Bismuth 0.086 2184Brass0.1

Strona 26

Dakota Ultrasonics 28 blank page

Strona 27 - APPENDIX B

PR-82 Sonic Tester 29 WARRANTY INFORMATION • Warranty Statement • Dakota Ultrasonics warrants the PR-82 against defects in materials and workmanshi

Strona 28

Dakota Ultrasonics 30 blank page

Strona 29

MATERIAL SAFETY DATA SHEETN/A = not applicable or not available (To comply with 29 CFR 1910.1200)SECTION 1 – PRODUCT IDENTIFICATIONProduct Name: SOUND

Strona 31 - APPENDIX C

Dakota Ultrasonics blank page

Strona 32

PR-82 Sonic Tester 1 INTRODUCTION The Dakota Ultrasonics model PR-82 is a precision Ultrasonic Micrometer. Based on the same operating principles a

Strona 33 - WARRANTY INFORMATION

Dakota Ultrasonics 2blank page

Strona 34

PR-82 Sonic Tester 3 OPERATION The PR-82 interacts with the operator through the membrane keypad and the LCD display. The functions of the various

Strona 35 - SONOTECH, INC

Dakota Ultrasonics 4The PRB-0 key is used to "zero" the PR-82 in much the same way that a mechanical micrometer is zeroed. If the tool is

Strona 36

PR-82 Sonic Tester 5 The UP arrow key has two functions. When the PR-82 is in calibration mode, this key is used to increase numeric values on the d